Precision Thermometer
For calibrating flat bed incubators and workstations
The ASL WIKA CTH7000 precision thermometer is a resistive type thermometer with an accuracy of ±0.02 compared to common used thermo couple types with accuracy of ±0.1.
The resolution of the ASL WIKA CTH7000 is 0.001°C needed for a correct calibration of your flat bed incubator.
• System accuracies better than 0.035°C over the full operating range.
• Stability: <0.005°C per year.
• Statisical analysis- average, max/min and standard deviation.
• Internal logging, software and USB communication
Availability
This product is available for sale in selected countries.
Product Specifications
Accuracy | ASL WIKA CTH7000 only: ±0.02°C (±20mK) |
Resolution | 0.001°C |
Probe types | Pt100 and Pt25 resistance thermometers plus NTC thermistors. Thermometer input2 x 5 pin industrial DIN socket with screw lock to connecting plug, accepting both normal and SMART probes |
Operating conditions | Full specification range: +15°C to +25°C Full operating range: 0°C to +40°C <80%RH non-condensing |
Dimensions (LxDXW) | 232x97x53 mm nominal |
Weight | 500 g (1.1 lbs) |
Order Codes
Order Number | Description |
---|---|
11010 | F100 Thermometer – 2 input |
11010P1 | F100 Thermometer, calibrated/1 probe – Calibrated with 1 probe |
11010P2 | F100 Thermometer, calibrated/2 probes – Calibrated with 2 probes |
11006 | STS Sensor – Temperature Sensor for validation |
22306 | Pin Sensor – F100 probe for incubator G185ST |